Chapter 4
Replace the O-Ring
Maintaining a Cold On-Column Injector for Standard Automatic Injections
Maintenance and Troubleshooting Manual
83
OPERATING SEQUENCE
Replace the O-Ring
NOTE
This sequence is used for the Automated Operation Cold On-Column Injector for TriPlus
Sampler.
Materials needed
•
Flat screwdriver
•
O-Ring (See # 5, Table 4-1 on page 81)
•
Tweezers
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
70
70 <
Initial time
1.0
Ramp 1
20.0
2. Scroll to
Temp
and press
OFF
.
3. Press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating, scroll
to
Pressure
and press
OFF
.
LEFT CARRIER (OCI)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
4. Remove the injector/detector cover.