Chapter 14
Replace the Interferential Filter
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
299
6. Loosen the knurled nut that fixes the
photomultiplier tube and remove it from the
detector body.
CAUTION
The photomultiplier tube could
damage if exposed to ambient
light with the excitation voltage
set to
On
. Make sure the GC
power supply has been
switched off before
disconnecting the tube from the
detector body.
7. Remove the interferential filter from its
housing, handling it very gently. Keep it using a
clean paper towel.
CAUTION
Filters are fragile. Pay attention
not to let the filter fall down and
damage.
Avoid touching the filter with
your fingers. If you see
fingertips on the filter, clean it
using GC-grade methanol and
air dry before remounting.