Chapter 5
Replace the Rotary Valve Seal
Maintaining a Cold On-Column Injector Direct Narrow-Bore
Maintenance and Troubleshooting Manual
109
28. Press
OVEN
, scroll to
Temp
and press
ON
.
29. Perform a leak check according to
Sequence on page 181.
30.
Open
Close
From the data system in use, enable the
automatic actuator to open and close the rotary
valve of the OC injector by using the direct
command
OC Actuator
. The actuator move
the OC valve lever as shown in the figure on the
right.