Chapter 17
Maintaining Detectors in Stack Configuration
Replace the Silver Seal (NPD on ECD)
380
Maintenance and Troubleshooting Manual
16. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, and reset
the ECD to the required operating conditions.
17. Press
AUX
.
18. Scroll to
Detector
and press
ENTER
.
19. In the
AUX DETECTOR (NPD)
table, set the
NPD detector to the required operating conditions.