Chapter 14
Maintaining a Flame Photometric Detector
FPD Troubleshooting
312
Maintenance and Troubleshooting Manual
Troubleshooting the FPD
Table 14-5.
FPD troubleshooting table
Symptom
Diagnosis
Remedy
High standing current
and noise
Detector conditioning not
completed
Increase detector conditioning time
Air flow rate set too high
Reduce air flow rate
Detector not light-tight
Tighten the chimney cap and the
knurled nut that fixes the
photomultiplier tube
lnterferential filter missing
Check the interferential filter has been
correctly installed
No standing current
Signal or power cables
incorrectly connected
Check connections
Faulty photomultiplier tube
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Unstable and
excessively noisy
baseline
Faulty photomultiplier tube
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
FPD temperature does
not reach the set point
Power cable not properly
connected
Check cable connection
Faulty heater
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.