Chapter 8
Maintaining a Purged Packed Column Injector
Replace the Septum
134
Maintenance and Troubleshooting Manual
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (PPKD)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Remove the septum using a non-metallic sharp
tool.