Chapter 9
Maintaining a PTV Injector
PTV Injector Maintenance
144
Maintenance and Troubleshooting Manual
PTV Injector Maintenance
The PTV injector will normally be serviced by Thermo Fisher Scientific
authorized technical personnel. In order to operate at peak performances, the
injector requires periodic maintenance from the user. This maintenance includes:
•
the replacement of the septum
•
the cleaning or replacement of the liner.
In addition, you will find information on how to adjust the quartz wool packing
inside the liner used in the Large Volume version of the PTV injector (PTVLV).
When replacing the septum
If you use the suggested syringe (P/N 365 020 01), the septum of your PTV
injector should be replaced after about 200 injections. The septum should be also
replaced every time a problem that is related to septum damage or wear occurs
(refer to Chapter 19,
10
Graphite ferrule for 0.2 mm ID column
290 134 89
Graphite ferrule for 0.25 mm ID column
290 134 88
Graphite ferrule for 0.32 mm ID column
290 134 87
Graphite ferrule for 0.53 mm ID column
290 134 86
11
Fixing nut for column
350 324 23
12
Liner removing tool
398 014 04
13
Screwdriver
205 026 03
14
Ceramic washer
344 011 00
14a
Clip for ceramic washer
Viton
®
O-ring for sintered liner
290 013 05
Kalrez
®
O-ring for sintered liner
290 037 02
Sealing cup for PTV liner (to be used with 290 013 05 or 290 037 02) 290 034 20
Table 9-1.
PTV Injector parts identification table (Continued)