Chapter 10
Performing a Leak Check (Capillary Column)
Ensuring Tightness
Maintenance and Troubleshooting Manual
181
OPERATING SEQUENCE
Performing a Leak Check (Capillary Column)
This procedure allows to check for leaks on the injector side only and it is suitable
only for capillary column.
Before starting this sequence, install the column into the injector port only,
leaving free the column end.
This Leak Check procedure can be easily performed by means of the provided
column-flow meter connector shown in Figure 10-1.
Flowmeter connection
fitting
Knurled nuts
Blind channel
for leak check
Open channel
for Col. Evaluation
Figure 10-1.
Column-flow Meter Connector
Materials needed:
•
Leak detector (Thermo Scientific GLD Pro, or equivalent)
•
Column-flow meter connector
The column-flow meter connector features two dedicated channels:
— A
blind channel
used to seal the column end when a
leak check
is
performed.
— An
open channel
used to measure the flow at the end of the column when
a
column evaluation
is performed (not used in this procedure).
1. Confirm that the carrier gas is on.
2. Carefully push the capillary column end into the
blind channel
of the
column-flow meter connector as shown in Figure 10-2.