6
•
Operation
UHPLC System
Operating Manual
Page 91
•
Never leave buffers, salt solutions, or aggressive solvents in the
system for a longer period without flow.
•
Note the special properties of the solvents, such as viscosity, boiling
point, or UV absorption.
•
If a charged aerosol detector is connected in the system flow path:
Observe the specific mobile phase guidelines for the detector in the
Vanquish Charged Aerosol Detector Operating Manual
.
TIP
In a Vanquish Core system, normal-phase (NP) compatible solvents
and additives may be used if the VC-pumps and the VC-autosamplers are
modified with the components from the Normal-Phase (NP) kit. Refer to
the
Operating Manuals
for the pumps and autosamplers.