8
•
Troubleshooting
Page 124
UHPLC System
Operating Manual
Post-column cooler does not reach set temperature
Possible Cause
Remedial Action
Post-column cooler not
installed correctly
Verify that the post-column cooler screws are tightened and the post-column
cooler is seated properly.
Autosampler does not reach set temperature
Possible Cause
Remedial Action
Ambient temperature
or humidity too high
Reduce the ambient temperature (for example, by ventilating the room), and/or
ambient humidity.
8.2.9
Power and Communication
Entire system has no power (all LED status bars unlit)
Possible Cause
Remedial Action
System power button
turned off
Turn on the system with the power button on the front left of the system base.
Single module has no power (LED status bar unlit)
Possible Cause
Remedial Action
Module power turned
off
Turn on the module with its main power switch.
Module not connected
to the mains
Connect the power cord.
Fuse has blown
Replace the fuses (refer to
Operating Manual
of the module).
Error in electronic
system
Contact Technical Support.
One or more modules cannot be turned off from the power button on the system base
Possible Cause
Remedial Action
System power
connection
interrupted
Check the system interlink connections between the system base and the system
modules, see
Summary of Contents for thermoscientific Vanquish UHPLC System
Page 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Page 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Page 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Page 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Page 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Page 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Page 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Page 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Page 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Page 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...