8
•
Troubleshooting
Page 122
UHPLC System
Operating Manual
Possible Cause
Remedial Action
UV/VIS lamp not stable
Allow the lamp and optics to warm up for at least 60 minutes. If the lamp is old, you
may need to replace the detector lamp (see
Detector Operating Manual
).
A new lamp may need burn-in time before the first analysis. For details, refer to the
Detector Operating Manual
.
Absorption of eluent
changes when gradient
is run
Absorbing additives may change the absorption spectrum, depending on the
solvent. Consider varying additive concentrations to level the drift.
Non-periodic baseline fluctuation, high noise
Possible Cause
Remedial Action
Pressure fluctuations
from pump
Purge the pump; check general function (refer to
Pump Operating Manual
).
Air trapped in the
system
Purge the system as necessary (refer to
Pump Operating Manual
).
UV/VIS lamp old or not
properly installed
Check that the lamp is correctly installed. Replace the lamp (refer to
Detector
Operating Manual
).
Improper reference
wavelength
The sample must not absorb in the range of the reference wavelength. Consider
using a method without reference wavelength.
Detector response time
too short
Select a suitable response time.
Improper wavelength
or optical bandwidth
Select an appropriate wavelength.
Select a higher bandwidth In particular with critical conditions (low absorption, few
light).
Flow cell not properly
installed
Check that the flow cell is installed properly (refer to
Detector Operating Manual
).
Detector defective
Contact Technical Support.
Optical transmission of
the flow cell too low
(Vanquish system with
VH-D10 detector only)
Check the lamp age and replace the lamp, if necessary. Use the
Cell Transmission
Check / Flow Cell Wash
button on the ePanel for the module and follow the
instructions. For this procedure, a diagnostic cell for the VH‑D10 is required. If you
do not have a diagnostic cell, compare the cell transmission (light intensity) values
of the flow cell before and after a wash procedure and check whether the cell
transmission (light intensity) has increased. If it has not increased, perform an
extended wash cycle and check whether the cell transmission (light intensity) has
increased.
Periodic baseline fluctuation, pulsation
Possible Cause
Remedial Action
Pressure fluctuations
from pump
Purge the pump; check general function (refer to
Pump Operating Manual
).
Air trapped in the
system
Purge the system (refer to
Pump Operating Manual
).
Improper reference
wavelength
The sample must not absorb in the range of the reference wavelength. Consider
using a method without reference wavelength.
Summary of Contents for thermoscientific Vanquish UHPLC System
Page 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Page 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Page 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Page 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Page 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Page 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Page 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Page 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Page 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Page 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...