8
•
Troubleshooting
UHPLC System
Operating Manual
Page 127
8.2.10.3
Intensity Test
The
Intensity Test
is a test to measure the intensity of the detector lamp
and the transmission performance ("intensity") of the optics. The test is
available for the diode array detector and for the variable wavelength
detector.
Follow these steps
1. Install the diagnostics cell in the detector.
2. Turn on the detector lamps (one or both) that you want to test.
3. Run the test from the Chromeleon software. Refer to the
Chromeleon Help
.
4. Depends on the test result:
Test result
Action
Passed
Reinstall the flow cell of your application.
Failed
1.
Diode Array Detector only:
Check the diagnostics
cell for indications of contamination that prevent the
light to pass the cell unhindered. Repeat the test with
a clean cell. If the test fails again, replace the lamp.
2. Replace the lamp. Refer to the
Operating Manual
for the detector.
3. If the test fails with the new lamp, the optics may
be defective. Contact Thermo Fisher Technical
Support.
8.2.10.4
Shutter Motor Test
The
Shutting Motor Test
checks the mechanical stability of the shutter
motor and filter paddle. The test is available for the variable wavelength
detector.
Follow these steps
1. Install the diagnostics cell in the detector.
2. Turn on the detector lamps (one or both).
3. Run the test from the Chromeleon software. Refer to the
Chromeleon Help
.
4. Depends on the test result:
Test result
Action
Passed
Reinstall the flow cell of your application.
Failed
Contact Thermo Fisher Scientific Technical Support.
Summary of Contents for thermoscientific Vanquish UHPLC System
Page 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Page 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Page 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Page 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Page 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Page 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Page 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Page 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Page 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Page 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...