Installation
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 69
In addition, observe the guidelines for proper connection of the flow cell
in
General Information and Guidelines (
.
Parts required
•
Inlet capillary
•
Detector waste line
For instructions on connecting the waste line, follow the steps in
Connecting the Detector Waste Line (
.
Tools required
For the detector waste line:
Tubing cutter (optional)
Preparations
TIP
For easier guiding of capillaries to the flow cell inlet, remove the
right door of the detector, see
.
1. Flush the system modules and capillaries upstream of the device to
waste before you connect the flow cell to the system flow path.
Refer to the
Vanquish System Operating Manual
.
2. On the flow cell, remove the plugs from the flow cell inlet and
outlet.
TIP
Store the plugs of the flow cell, for example in the flow cell
packaging, to have them easily available when storing or shipping the
flow cell.
3. Install the partition panel plugs (if not present yet). See
Follow these steps
1. Connect the inlet capillary to the flow cell.
2. Connect the detector waste line to the flow cell.
See also
2
Connecting the Inlet Capillary (
2
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...