5
•
Installation
Page 60
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
TIP
Store the flow cell in a clean place for later use in diagnostic
procedures or for transporting the device.
5.7.2
Installing a Flow Cell without Connection Unit
Parts required
Flow cell, as required by the application
Preparations
1. Remove the diagnostic cell. See
2. Unpack the flow cell.
Follow these steps
1. Push the flow cell into the flow cell opening.
2. Tighten the two flow cell screws hand-tight.
3. Set up the flow connections to the flow cell, see
4. Perform wavelength validation, see
.
TIP
If the wavelength validation fails, perform a calibration of
wavelengths.
5.7.3
Installing a Flow Cell with Connection Unit
Parts required
Flow cell, as required by the application
Preparations
1. Remove the diagnostic cell. See
2. Unpack the flow cell.
Follow these steps
1. Push the flow cell partly into the flow cell opening.
2. Place the connection unit on the connection unit socket. Take care
that the locating pin is placed in the slot.
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...