![Thermo Scientific Thermo Scientific Vanquish VC-D40 Operating Manual Download Page 147](http://html.mh-extra.com/html/thermo-scientific/thermo-scientific-vanquish-vc-d40/thermo-scientific-vanquish-vc-d40_operating-manual_1105377147.webp)
8
•
Troubleshooting
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 147
Follow the steps depending on the location of the leak:
Location of the leak
Steps
Flow cell inlet and/or outlet
1. Tighten the connection where liquid is
visible.
2. If the connection seems tight but is still
leaking, remove the connection/fitting
and check for damage.
3. If necessary, replace the inlet capillary
or waste line.
Drain outlet of the flow cell (see figure),
or connection from flow cell to
connection unit
1. Remove the flow cell from the detector
and inspect the flow cell for signs of
leakage.
2. If signs of leakage are present at the
flow cell, replace the flow cell.
3. With a cloth or tissue, thoroughly absorb all liquid that has collected
in the leak tray and under the leak sensor. Be careful not to bend the
sensor.
4. Allow the sensor to adjust to the ambient temperature for a few
minutes.
5. If leakage is no longer reported, you can resume operation.
See also
2
Installing a Flow Cell without Connection Unit (
2
Installing a Flow Cell with Connection Unit (
2
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...