Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Figure 25: Screws on the lamp house cover
3. Remove the lamp house cover.
CAUTION—Hot surface
The lamp may become hot. Touching a hot lamp might cause burns.
•
Touch the lamp briefly and carefully to find out if it is hot before you
remove the lamp.
•
If it is hot, wait until the lamp has cooled down.
4. Push the clip on the lamp connector and disconnect the connector
from the lamp socket on the device.
5. With your hands, loosen the two screws that attach the lamp.
6. Pull out the lamp.
7. Align the new lamp with the locating pin.
When the lamp is in the correct position, push the lamp gently into
the lamp house. The lamp socket must be in a level position with the
lamp house.
8. When the lamp is properly seated, tighten the two screws to attach
the lamp to the lamp house.
9. Connect the lamp connector.
10. Mind the routing of the lamp cables.
Position the lamp cables as indicated in
to prevent the cable from being pinched under the
lamp house cover.
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...