8
•
Troubleshooting
Page 142
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Message and Code
Description and Remedial Action
Code 7049 Wavelength calibration failed
– check and flush flow cell.
Check the flow cell (see
the message still appears, proceed as follows:
Check the lamps: If a lamp was newly installed, ensure that it is
installed properly. If a lamp is old (counting, for example, more
than 2000 operating hours), replace the lamp (see
Code 7051 Bad flow cell position.
The flow cell is not correctly positioned inside the flow cell
opening.
Check the flow cell installation (see
Check the flow cell contact pins for proper alignment inside the
flow cell opening.
Code 7056 Holmium validation failed –
wavelength outside limits.
Performing a Wavelength Validation
). If the message still appears,
proceed as follows:
Perform a calibration and repeat the validation again.
Remove and install the flow cell. Flush the flow cell with pure
HPLC-grade water with a flow rate of > 1 mL/min for several
minutes (Nano flow cells: Flush with pure HPLC-grade water with
the maximum possible flow rate for the respective range for
several minutes).
Install a different flow cell or a diagnostic cell (see
or
Installing a Diagnostic Cell (
Perform a validation. Perform a calibration and repeat the
validation again. If the message disappears with a different cell,
the flow cell that was previously installed may be clogged.
If the error message persists, contact Technical Support.
Code 7057 Holmium validation failed –
specified wavelength not found.
Repeat the validation. If the message still appears, proceed as
follows:
Perform a calibration and repeat the validation again.
Remove and install the flow cell. Flush the flow cell with pure
HPLC-grade water with a flow rate of > 1 mL/min for several
minutes (Nano flow cells: Flush with pure HPLC-grade water with
the maximum possible flow rate for the respective range for
several minutes).
Check the lamp: If the lamp was newly installed, ensure that it is
installed properly. If the lamp is old (counting, for example, more
than 2000 operating hours), replace the lamp. See
Install a different flow cell (see
Perform a validation. Perform a calibration and repeat the
validation again. If the message disappears with a different cell,
the flow cell that was previously installed may be clogged.
Code 7059 Calibration failed – values
outside specs.
Remove and install the flow cell.
Flush the flow cell with pure HPLC-grade water with a flow rate of
>1 mL/min for several minutes (Nano flow cells: Flush with pure
HPLC-grade water with the maximum possible flow rate for the
respective range for several minutes).
If the error message persists, install a different flow cell or a
diagnostic cell (see
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...