3
•
Device Overview
Page 32
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
3.1
Detector Features
The detector comprises the following main features:
•
The detector is a dual-beam, variable wavelength photometer with
one measurement and one internal reference beam.
•
Two different light sources provide a wavelength detection range
from 190 nm to 750 nm (VC-D40) and from 190 nm to 900 nm (VF-
D40):
¨
A deuterium lamp (referred to as UV lamp) for ultraviolet and
visible detection and
¨
A tungsten lamp (referred to as VIS lamp) for visible and near-
infrared detection (optional for VC-D40)
•
Internal validation of the wavelength accuracy with holmium oxide
filter
•
Multi-Wavelength Operation for additional operation functions,
measuring simultaneously at up to two (VC-D40) or four (VF-D40)
different wavelengths
•
Data collection rates of up to 125 Hz (VC-D40) and up to 250 Hz (VF-
D40) under Chromeleon 7.2 software or later, in single-channel
mode
•
Spectra measurements
•
Two optical filters for eliminating higher order light
To ensure optimal measurement results over the entire wavelength
range, the optical filters automatically move into the light path (also
in multi-wavelength operation).
•
Automatic wavelength calibration after power-up and when the
lamp configuration was changed.
•
A shutter (motorized filter paddle) to move into the light path before
the flow cell
The shutter can move into the following positions:
¨
Open position for data acquisition
¨
Closed (dark) position for protection of the flow cell and for dark
signal measurements
¨
Holmium oxide filter position for validation of wavelength
accuracy
•
Availability of flow cells for different fields of applications (for
details, see
)
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...