5
•
Installation
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 61
4
2
1
3
Figure 11: Installing a flow cell with connection unit
No.
Description
1
Connection unit
2
Locating pin
3
Connection unit socket
4
Connection unit screw (knurl)
3. Tighten the connection unit screw hand-tight.
4. Push the flow cell completely into the flow cell opening.
5. Tighten the two flow cell screws hand-tight.
6. Set up the flow connections on the flow cell, see
.
7. Perform wavelength validation, see
.
TIP
If the wavelength validation fails, perform a calibration of
wavelengths.