Maintenance and Service
Page 114
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
7.6
Flow Cell
This section describes cleaning, replacement, storage and shipment of
flow cells.
To store or ship the flow cell, follow these steps:
1. Flush the flow cell with isopropanol.
2. Remove the flow cell from the detector. See
.
3. Install the plugs on the flow cell inlet (IN) and outlet (OUT), and
place the flow cell in its packaging.
To exchange the flow cell, follow these steps:
1. Flush the flow cell with isopropanol.
2. Remove the flow cell from the detector. See
.
3. Install a new flow cell without connection unit to the detector. See
Installing a Flow Cell without Connection Unit (
– or –
4. Install a new flow cell with connection unit to the detector. See
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...