Maintenance and Service
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual
Page 107
7.5
Lamps
This section describes the replacement and installation of the
•
Replacing the UV Lamp (Deuterium Lamp)
•
Replacing the VIS Lamp (Tungsten Lamp)
NOTICE—Damage to the device
Usage of lamps not supplied or approved by Thermo Fisher Scientific for
this detector may result in inferior performance, detector malfunction or
may damage the detector. Detector guaranty may be voided.
Use only lamps supplied or approved by Thermo Fisher Scientific for this
detector.
7.5.1
Replacing the UV Lamp (Deuterium Lamp)
When
•
The lamp is defective
•
After a defined amount of operating hours of the lamp as a
preventive measure
•
The lamp intensity is too low for demanding applications
•
The lamp intensity fluctuates causing artifacts in the baseline
Parts required
•
UV lamp
•
Optionally:
¨
Isopropanol
¨
Lint-free tissue
Preparations
1. Turn off the UV lamp.
2. Turn off the device.
Summary of Contents for Thermo Scientific Vanquish VC-D40
Page 4: ...Contacting Us Page 4 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 10: ...Contents Page 10 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 98: ...6 Operation Page 98 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 148: ...8 Troubleshooting Page 148 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 154: ...9 Specifications Page 154 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 168: ...11 Appendix Page 168 Variable Wavelength Detectors VC D40 VF D40 Operating Manual ...
Page 173: ...Index Variable Wavelength Detectors VC D40 VF D40 Operating Manual Page 173 ...