Chapter 4
|
Water Analysis Test Menu
34
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Loading Test Methods from USB Stick to AquaMate Instrument
1.
Open the AquaMate USB memory stick using a computer.
2.
Copy the desired test methods from the Orion, Merck and CHEMetrics folders to the
Thermo folder.
3.
Remove the USB memory stick from the computer and insert it into the USB port on the
front of the AquaMate instrument.
4.
When the AquaMate instrument is first powered on, it will display the initial measurement
screen. Press the test key to access the Test Types menu.
5.
Press the ▲or ▼ key to highlight Water Analysis and press the enter key.
6.
Press the Stored Tests function key.