Chapter 8
|
Absorbance, % Transmittance and Concentration Measurements
154
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Multiwavelength
The Multiwavelength application obtains multiple fixed-wavelength measurements. It is a fast
alternative to scanning if the wavelengths of interest are well known. This section covers:
•
Setting Up Test Parameters and Adding Wavelengths and Factors
•
Taking Measurements
•
Recalling a Test
Setting the Parameters for Multiwavelength Method
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Multiwavelength and press the enter key.
3.
Highlight and change the displayed test parameters, including Test Name, Measurement
Mode and Sample Positioner.
a.
Some parameters only appear when the concentration mode is selected, while others
appear regardless of the selected measurement mode. See the
for more information on the test parameter settings.
4.
Press the Set nms function key to add the wavelengths and factors. Factors can only be
added when the measurement mode is set to Conc/Factor.