Chapter 7
|
Wavelength Scanning Test Menu
Thermo Scientific
Orion AquaMate Spectrophotometer User Manual
|
133
Setting the Parameters for a Scan
1.
Press the test key to access the Test Types menu.
2.
Press the ▲or ▼ key to highlight Scanning and press the enter key.
3.
Highlight and change the displayed test parameters, including Test Name, Measurement
Mode, Start Wavelength, Stop Wavelength, Sample Positioner, Scan Speed and Interval.
a.
See the
section for more information on the test parameter settings.
i.
Use the
to set a baseline expiration time.
ii.
If Auto Save Data is set On, enter a Data File Name before accessing the Run
Test or Measure Samples functions.
b.
Sample Positioner Setting for the Vial Holder Carousel
information on selecting the appropriate Sample Positioner setting.
i.
Scan tests only measure one sample at a time, so Auto 6 and Auto 3 Sample
Positioner settings are not fully functional.
4.
Press the Save Test function key and follow the onscreen prompts to save the test.