Chapter 10
|
Maintenance
182
|
Orion AquaMate Spectrophotometer User Manual
Thermo Scientific
Changing the Fuse
The fuse is located in the power entry module located at the center of the back panel of the
instrument.
•
120 VAC, 2.5 A, Slo Blo
•
240 VAC, 1.25 A, Slo Blo (2 required)
Important: The instrument fuse must be replaced with the same type and rate.
Important: If the fuse fails repeatedly, it may indicate a serious problem with the instrument.
Contact technical support as soon as possible.
To change the fuse:
1.
Turn off and unplug the instrument from the wall outlet or power strip.
2.
Position the instrument so you can access the power entry module on the back of the
instrument.
3.
Remove the power cord.
4.
Insert a flat-blade screwdriver into the notch on the fuse cover and pry off the cover.