Troubleshooting
73
Error message
Symptoms
Possible causes and corrective measures
E-06
Rotor stops without de-
celeration to standstill.
Instrument cannot be
operated.
Communication error between key panel and CPU
Switch the instrument off and on again. If the error persists, con-
tact customer service.
E-07
The rotor is decelerated
to a standstill. The lid can
be opened.
Over temperature in the centrifuge chamber
Display > 51
°C or measured temperature > 70
°C.
(Cooling machine possibly defective.)
E-08
Rotor stops without de-
celeration to standstill.
Instrument cannot be
operated.
Over voltage at frequency converter
.
Mains voltage outside the tolerance. Deceleration resistance de-
fective. If necessary, request customer service.
E-10
Self-test after switching
on the centrifuge.
NV-RAM; error in program memory
Switch the instrument off and on again. If the error persists, con-
tact customers service.
E-12
Rotor stops without de-
celeration to standstill.
Instrument cannot be
operated.
Temperature measurement error.
Switch the instrument off and on again. If the error persists, con-
tact customer service.