Troubleshooting
68
By joltingly pulling the attached cord the mechani-
cal lid unlocking mechanism is activated. The lid
will open and you can remove your samples.
4. Afterwards push the cord back into the instrument
and close the opening by the plug again.
Once the power failure has been eliminated, you can
connect the instrument to the mains and switch it on.
Following the self test of the centrifuge, the lid may be
closed and locked with the motor.