Operation
58
Centrifuging with a program
After closing the centrifuge lid, recall the
desired program memory number using
the program selection key and actuate the
start key
.
If the rotor is started with a program the speed or
RCF set value of which exceeds the permissible one
for the inserted rotor or the RCF/set value of which is
below the rotor-specific minimum, the display in the
program display section is changed to [ ] after the
identification of the rotor.
“Quick Run”
For short-term operation, the
Cryofuge
®
5500i GMP
is
equipped with a “Quick Run” function
.
Short-term centrifugation is started by pressing the
“quick run” key
continuously; it stops as soon as
the key is released.
In this mode the centrifuge accelerates and deceler-
ates with full power. The set speed or RCF is ignored
in this case.
During acceleration the time is counted forward in
seconds. The display stays until the centrifuge lid is
opened.
Depending on the rotor installed, the
centrifuge accelerates to the maxi-
mum speed!
Check carefully whether you have
to maintain a specific speed limit
for your application.