Troubleshooting
67
Troubleshooting
Emergency lid release
In case of a power failure the lid cannot be opened
using the electrical lid release
.
To permit unloading in
this case, the centrifuge is equipped with an emer-
gency lid release. However, you may use this system
in case of emergency only.
Proceed as follows:
1. Make sure the rotor is at a standstill
(sight glass in the lid).
2. Pull mains plug.
3. On the right- and left-hand side of the front screen
there is a plastic plug, which you can remove with
a screwdriver or a knife.
Emergency release
Rotor can run at high speed! Touch-
ing it may cause severe injuries
!
Always wait several minutes until
the rotor has come to a complete
standstill without deceleration.
Without power supply the brake is
out of operation, and deceleration
takes much longer than normal!
During a power failure it is impossible
to lock the lid once the emergency lid
release has been actuated
!
Never decelerate the rotor using your
hands or tools
!