Operation
56
Stopping the centrifuge
At limited run time
Normally the run time has been set manually and all
you have to do is wait until the centrifuge terminates
the run automatically at the end of the set time.
As soon as the speed reaches zero, the display reads
“end”. You can now open the centrifuge by pressing
the “open lid” key
and remove the centrifuged
material.
If the lid has not been fully lifted out of the lid lock, the
message “lift lid” appears (manual lift of the lid is nec-
essary).
You can manually stop the centrifuge at any time by
actuating the “stop” key .
At this point the remaining run time is displayed.
At continuous operation
If you have chosen continuous operation, you must
stop the centrifuge manually. Press the “stop” key
in the control panel. The centrifuge will be deceler-
ated with the preset deceleration profile.
The display reads “end”, and you can open the lid by
pressing the “open lid” key
and remove the centri-
fuged material.
Temperature control during standby
The exact regulation will be active after identification
of the rotor. This is the case after a centrifugation run
exceeding 300 rpm. At standby the speed display
reads “end”.
If the rotor has not been identified (lid has been
closed and the “start” key
has not yet been actu-
ated, speed display shows “0” with flashing point), the
temperature regulation inhibits the freezing of the
sample only. An exact regulation cannot be guaran-
teed.