Control panel of the Cryofuge
®
5500i GMP
Displays
Program selection
Program selection key: stepwise recall of programs
Switch key:
open/shut program storage
Acceleration profiles
(1= slow ........ 9= fast)
display:
acceleration profile last set 1- 9
Deceleration profiles
(1= slow ........ 9= fast)
display:
deceleration profile last set 1- 9
Speed / RCF
run:
actual value of speed or RCF after actuating
switch key
end:
“End”
lid open:
“OPEN”
“Lift Lid” (if lid is not automatically lifted off)
Keys
start :
normal start of the centrifuge
stop :
manual stop of a run
open lid:
open lid (possible only with the instrument
switched on)
Quick run:
short-term operation of the centrifuge as long as
key remains pressed
switching
rpm/RCF:
switching between rpm and RCF display
Bucket set:
setting of the bucket number
Pretemp :
pre-temp-function
“set” keys:
stepwise increase/decrease of set values
Short pressing of any of these keys: switching from actual to set
value, signaled by flashing display
(before start)
lid closed:
“0” with flashing point
(rotor not yet identified)
error code:
will flash in display
Run time
time selection: - remaining run time up to 0
continuous
- run time passed
operation (hold) (in minutes and seconds)
“quick run”:
- run time passed
(in minutes and seconds)
Temperature
run:
actual sample temperature in °C
(in temperature equilibrium)
Troubleshooting see chapter “Troubleshooting”):
E-00: motor blocked
E-03: speed measurement
E-08: over voltage; over temperature in the electronics
E-14: no rotor or rotor identification not possible
E-17: lid cannot be opened
E-23: under voltage; over temperature in the motor or
excessive pressure in the refrigerator
rotor: set speed higher than highest permissible speed
of the rotor
bAL : imbalance
Lid:
lid burst open during run or open