78
4. When the scope has finished
acquiring segments, press the
Segments Stop
key from the
bottom menu.
Alternatively, press the
Run/Stop
key again.
The Stop Indicator will be shown
when in the Stop mode.
or
Run/Stop
The scope is now ready to navigate or analyze the
acquired segments.
4-2-4. Navigate Segmented Memory
Background
After the segmented memory acquisitions have
been captured you can navigate through each
segment one at a time.
Operation
1. Press
Select Segments
from the
bottom menu. This key will be
available in the Stop mode.
2. To navigate to the segment of
interest, press
Current Seg
from
the side menu and use the Variable
knob to scroll to the segment of
interest.
Alternatively, the Set to Minimum
and Set to Maximum keys can be
used to jump to the first and last
segment respectively.
3. The position in time of the selected
segment relative to the time of the
first segment is shown in the
Segments Time
key.
4-2-5. Segment Measurement
Background
The Segmented memory function can be used in
conjunction with the automatic measurements in
the Measurement menu. Please note that digital
channels are not supported for measurement using
segments.
Display All The Display All function will show
all the acquired segments