105
4-6-9. Pulse Runt trigger
Panel Operation 1. Press the trigger
Menu
key.
Menu
2. Press the
Type
key from the lower
bezel menu.
3. Select
Others
→
Pulse
Runt
from the side menu.
The Pulse and Runt
indicator appears at the
bottom of the display.
From left: Polarity, source, high/low threshold,
threshold level, coupling
4. Press
Source
from the lower menu.
Use the side menu to select a source.
Range
Channel 1 ~ 4
5. Press
Polarity
to toggle the polarity.
Range
Rising edge, falling edge, either.
6. Press
When
from the lower menu.
Then use the side menu to select the condition
and width.
Condition
> , < , = , ≠
Width
10ns ~ 10s
7. Press
Threshold
from the lower
bezel to edit the threshold for the
upper and lower threshold.
8. Use the side menu to set the upper
threshold.
Range
-XXV~XXV
Set to TTL 1.4V ,Set to ECL -1.3V
9. Use the side menu to set the lower
threshold.
Range
-XXV~XXV
Set to TTL 1.4V
Set to ECL -1.3V