175
Autoset does not catch the signal well.
The Autoset function cannot catch signals under 30mV or 20Hz.
Please use the manual operation. See page 45 for Autoset
details.
I can’t save files to the internal memory.
If a USB stick is inserted into one of the USB slots and you wish
to save to the scope internal memory, press the Utilities key
and set the file path to internal memory.
The display image printout is too dark on the background.
Use the Ink Saver function which reverses the background color.
For details, see page 148.
The date and time settings are not correct.
For date and time setting details, please see page 115. If it
does not help, the internal battery controlling the clock might be
worn out. Contact your dealer.
The accuracy does not match the specification.
Make sure the device is powered On for at least 30 minutes,
20°C~+30°C. This is necessary to stabilize the unit to
match the specification.
For more information, contact your local dealer or TEXIO
TECHNOLOGY at www.texio.co.jp /