
5 First steps
79
Pos: 83 /TD/Erste Schritte/testo Saveris/06 ****Probelauf/02 Alarmkette prüfen @ 0\mod_1189157374281_79.docx @ 4064 @ 2 @ 1
5.12.3.
Checking alarm chain
You can trigger an alarm and control the sending of the messages
without having to bring about a real alarm situation.
>
In the navigation area, click on
Alarm management
.
- In the data window, the
Basic setting
and
Alarm overview
submenus as well as the probes connected in the Saveris base
are shown.
1.
Click on
Basic setting
.
- The basic setting for the alarms is shown in the display range.
2.
Click on the
[SMS test report]
button.
- A test report is sent to the mobile phone number entered.
Pos: 84 /TD/Erste Schritte/testo Saveris/07 **** Montage der Hardware/00 Hardware montieren @ 0\mod_1189157579984_79.docx @ 4124 @ 3 @ 1
5.13.
Mounting the hardware
The notes in chapter "Ensure safety" must absolutely be
observed when mounting the Saveris components.
Do not mount the Saveris base and the probes at their
locations until the measuring system functions as
expected.
At sub-zero temperatures, the Li-ion rechargeable
batteries in the base and in the Ethernet probes are not
charged. Avoid assembly sites for these components
with long-lasting, prevalent sub-zero temperatures.
After mounting, perform another test run of the system;
also see "Performing the test run".