7 Maintaining the product
151
◦
For V1.x systems: software 4.2 SP3, base and radio probe
V1.90, router, converter and extender V2.59, Ethernet probe
V1.47
◦
For V2.x systems: software 4.2 SP3, base and radio probe
V2.59, router, converter and extender V2.59, Ethernet probe
V1.47
• All outstanding alarms must be acknowledged by you before a
firmware system update.
• ATTENTION: During the firmware update process, never
disconnect the power supply/network supply and the PC
connection.
• During the update, measurements, data storage and data
communication are continued so there is no data loss.
• Configuration changes during the firmware system update
should not be carried out to ensure that the process is not
slowed.
• As there is heavy radio traffic during the update, alarms relating
to interrupted radio links may occur temporarily.
• The reboot phase once a probe has been updated takes
approx. 1-2 minutes. During this brief time, no measuring
values can be recorded.
Carrying out an update
Please note that the entire update process may take
several hours. During this time, no configuration changes or
network/power supply interruptions should occur.
1. Unpack the firmware file folder and save it to your PC.
2. Open
Saveris Startup Wizard
and click in the
Projects
tab on
System Update
.
- A Windows® Explorer window
Search folder
is opened.