Question
Possible causes / solution
The converter does not
transfer any data to the
base.
The cable connection to the converter
is faulty.
> Remove the power supply and
check whether the Ethernet cable
is connected correctly.
> Reconnect the power supply.
- The converter checks its
configuration and if it detects an
error, it resets all values to the
factory settings.
Pos: 192 /TD/Tipps und Hilfe/Fragen und Antworten/testo Saveris/8.2 FAQ Alarmmeldungen Saveris Base @ 12\mod_1339687134056_79.docx @ 130826 @ 2 @ 1
8.2.
Saveris base alarm messages
Alarm message
Possible causes/solution
L_CommUp
L_CommApp
Error during USB or Ethernet
initialisation.
> Disconnect all links to the base.
> Reconnect all links.
> Restart the base.
L_GSM
L_GSMMenue
Error during GSM modem
initialisation.
> Check the GSM module's battery
power.
> Restart the base.
L_RF2010Server
L_RF2010IO
L_RF2010MemPool
L_RF2010StreamRip
L_UDPRF2010
Error during radio module initialisation.
Reboot the base. If the problem
persists, contact the Service team.
> Restart the base.
If the problem persists, please contact
testo Service.
L_UIPrio
L_DispDrvUI,
Error loading the UI/display. Reboot
the base.
> Restart the base.