It is also possible to pass again and to move to the opposite direction again in order to
define a new inflection point, which will be compared to the first one. This step can be
repeated as often as you want until the pressure on the probe is stopped.
7. The result is automatically displayed on the screen.
12.13 Bore
measurement,
dynamic mode
(MH)
Bores or axes cannot be measured in mode
ST1
with a manual MICRO-HITE, as it
requires probing in two directions. Therefore, mode
ST2
is required.
Once in mode
ST2
, it is important to verify that the option of probing with two contacts
has been correctly selected (for further details, see this
). Then proceed as
explained
for each upper and lower culmination point.
12.14 Culmination
point (MH+M)
On the MICRO-HITE+M, the type of measurement of the culmination point is intrinsically
defined by the selected keyboard action. The software knows the nature of the point to be
measured from the selected process:
Minimum internal culmination point
Maximum internal culmination point
Minimum external culmination point
Maximum external culmination point
Bore
Axis
The dynamic mode is the only mode defining the process to determine
culmination points with a motorised height gauge. It is the workpiece