![TESA MICRO-HITE Instruction Manual Download Page 69](http://html.mh-extra.com/html/tesa/micro-hite/micro-hite_instruction-manual_1090543069.webp)
Instruction manual for MH & MH+M
69
Note that a single probing block is created with the icon
(or
).
6. Proceed the same way for the next measurements, regardless of the probing direction.
12.7 Automatic single
probing (MH+M)
The automatic single probing process is completely automatic, once the user has chosen
an action from the control panel keyboard or via the displacement handle.
Process
1. Position the probe close to the zone to be measured using the handle on the measuring
carriage or the rotary control handle.
2. Make sure that there is no element between the probe and the zone to be measured
that could impede the displacement of the probe.
3. Start the measurement by activating the key
,
or start the measurement directly
using the rotary control handle (chapter 2.5).
4. As soon as the probe gets in contact with the point to be measured, the system ensures
a constant probing force. After a stabilisation period, the measurement will be saved
and then displayed on the control panel. The LED lights up briefly at the same time as
the beep that indicates the end of the measurement.