LabMaster 10 Zi Oscilloscopes
46
LM10Zi-GSM-E Rev A
Basic Controls
Front Panel Control Groupings
Miscellaneous Controls / WaveStream Indicator
Help
- Opens the Teledyne LeCroy Online Assistant. If the second
monitor is installed, the online help opens on the second
monitor.
Default Setup
- Resets the oscilloscope's settings to the default
configuration. Corresponds with screen menu selection:
File →
Recall Setup → Recall Default Setup...
. For a list of default
settings, see
Save/Recall → Saving and Recalling Setups
(on page
173).
Auto Setup
- Opens the
Auto Setup...
flyout menu.
Press the
Auto Setup...
button on the flyout menu to perform a
full auto setup. Press a
Channel Find Scale
button on the flyout
menu to perform a quick auto setup for that channel only. Press
the
A
UTO
S
ETUP
...
front panel button twice to perform the last
selection from the
Auto Setup...
flyout menu (the default is to
perform a full auto setup).
If Auto Setup is run when no channels are turned on, all channels
are affected. When more than one channel is turned on, the first
channel with a signal applied to it is automatically set up for edge
triggering.