LabMaster 10 Zi Oscilloscopes
106
LM10Zi-GSM-E Rev A
information on using the touch screen controls can be found in
the
Dialog Area
.
Provide a voltage level (in millivolts).
NOTE
: Once specified,
Trigger Level
and
Coupling
are the only
parameters remaining unchanged as you switch from trigger mode to
trigger mode for each trigger source.
Optimize for HF
The Optimize for HF checkbox can be used on an
Edge
trigger to reject
high or low frequencies.
The checkbox is marked by default; meaning, the instrument is
optimized for high-frequency waveforms.
NOTE
: If you are measuring a waveform that is 10 MHz or slower, be
sure to unmark the checkbox to avoid triggering on an incorrect slope.
Width Condition
Width
triggers (and other triggers equipped with conditional logic) have
conditional logic settings as follows.
Your
Width Condition
can be specified as
Less Than
,
Greater Than
,
In
Range
and
Out Of Range
.