LabMaster 10 Zi Oscilloscopes
182
LM10Zi-GSM-E Rev A
Touch-Screen Calibration
button launches a sequence of
calibration screens. These screens instruct you to touch a series
of displayed crosses in order to improve precision and accuracy
when using your touch screen. The calibration procedure times
out if no cross is touched for 10 seconds.
Service
button launches a section of the application reserved for
qualified Teledyne LeCroy personnel. An access code is required
to enter this section.
Status
The Status dialog displays information about your instrument, including
model number
,
serial number
,
firmware version
. It also shows your
activated hardware and software options.
Adding Software Options
Newly purchased software options are enabled on the oscilloscope
using a key code supplied by Teledyne LeCroy.
1.
From the menu bar, choose
Utilities > Utilities Setup
.
2.
Touch the
Options
tab.
3.
Touch the
Add Key
button.
4.
Enter the key code.
5.
Touch
Activate Key Now
.
6.
Restart the oscilloscope.
Remote Control
Use the
Remote
dialog to setup remote communication. You can select
a network communication protocol, establish network connections, and
configure the Remote Control Assistant log from the
Remote
dialog. The
choice of communication protocols is limited to TCPIP, GPIB, and
USB488 (on specific models).