LabMaster 10 Zi Oscilloscopes
94
LM10Zi-GSM-E Rev A
3.
Connect the Fast Edge output on the front of the Master Control
Module to Channel 1, then on the Wizard touch
Next
.
4.
When prompted, connected the Fast Edge output (using the
same cable as used to connect to Channel 1) to Channel 5, then
touch
Next
again.
5.
Continue following on-screen prompts for as many additional
modules there are.
6.
When completed, touch
Finish
to apply and store all corrections
to the calibration folder.