Getting Started Guide
LM10Zi-GSM-E Rev A
111
Training TriggerScan
The TriggerScan Trainer inspects a currently acquired waveform and
automatically builds a list of common trigger setups used to find rare
events.
NOTE
: Run the Trainer if you want to change the trigger types or if you
change the channel or signal. You must acquire and display at least 3
cycles of a signal before running the Trainer.
1.
Touch
Trigger → Trigger Setup...
from the menu bar.
2.
On the
Trigger
dialog, click the
TriggerScan
tab.
3.
Touch the
Trainer
button and select a source channel from the
pop-up. Also choose the types of triggers the Trainer should use
to train the system and then touch the
Start Training
button. The
training begins. When it is complete, a list of smart trigger setups
is displayed in the Trigger List.