6.6.1.
Analog Output Range Selection and Offset Adjustment
The final step in configuring the analyzer’s three analog output channels is to set
the electronic signal type and range of each channel. This consists of selecting a
type, voltage or current (if an optional current output driver has been installed),
and a signal level that matches the input requirements of the recording device
attached to the channel. A bipolar offset can be also added to the signal if
required.
To select an output signal type (DC Voltage or current) and level for one output
channel, activate the
ANALOG I/O CONFIGURATION
MENU
, then press:
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO OUTPUT RANGE:
5V
0.1V 1V 5V
10V
CURR
ENTR
EXIT
These keys set
the signal level
and type of the
selected channel
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the
analog output channel to be
configured. Press
EDIT
to
continue
Pressing
ENTR
records the new setting
and returns to the previous menu.
Pressing
EXIT
ignores the new setting and
returns to the previous menu.
DIAG AIO OUTPUT RANGE:
10V
0.1V 1V 5V
10V
CURR
ENTR EXIT