Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B
101
6.7.3.3.
Selecting for Auto/Manual Analog Output
Calibration
The analog outputs configured for voltage mode can be calibrated either
automatically or manually. In its default mode the instrument is configured for
automatic calibration. Manual calibration should be used for the 0.1V range or in
cases where the outputs must be closely matched to the characteristics of recording
device. Outputs configured for automatic calibration can be calibrated as a group or
individually.
To select manual output calibration for a particular channel press the following:
DIAG AIO AOUT AUTO CAL: ON
ON ENTR EXIT
Exit to Return
to the main
Sample Display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
Toggles the Auto
Cal Mode
ON/ OFF for this
Analog Output
channel only.
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A4
Then Press
EDIT
to continue
Pressing
ENTR
records the new setting
and returns to the previous menu
Pressing
EXIT
ignores the new setting and
returns to the previous menu
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET>
EDIT
EXIT