Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
130
P/N 04316 Rev: B
6.9.9.
Manually Configuring the Network IP Addresses
There are several circumstances when you may need to manually configure the
interface settings of the analyzer’s Ethernet card. The
INET
sub-menu may also be
used to edit the Ethernet card’s configuration properties
•
Your LAN is not running a DHCP software package,
•
The DHCP software is unable to initialize the analyzer’s interface;
•
You wish to program the interface with a specific set of IP addresses that
may not be the ones automatically chosen by DHCP.
Editing the Ethernet Interface properties is a two step process.
STEP 1: Turn DHCP
OFF
: While DHCP is turned
ON
, the ability to manually set
INSTRUMENT IP
,
GATEWAY IP
and
SUBNET MASK
is disabled
ENTR
accept
new settings
EXIT
ignores
new settings
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE RANGE = 50.0 PPM CO =XX.X
< TST TST > CAL
SETUP
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X
DHCP: ON
ON
ENTR EXIT
SETUP X.X
DHCP: ON
<SET SET>
EDIT
EXIT
SETUP X.X
DHCP: ON
OFF
ENTR
EXIT
Continue with editing of Ethernet interface
properties (see Step 2, below).