Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
102
P/N 04316 Rev: B
6.7.3.4.
Analog Output AutoCal
Outputs configured for automatic calibration can be calibrated either individually or
as a group. To automatically calibrate an individual channel press the following:
DIAG AIO
CONC_OUT_2 CALIBRATED:
YES
<SET CAL
EXIT
EXIT
to Return
to the main
Sample Display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
<
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET>
EDIT
EXI T
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
<SET
CAL
EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
Press
SET>
to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A4
Then Press
EDIT
to continue