EPA Protocol Calibration
Model 400E Ozone Analyzer
Instruction Manual
202
P/N 04316 Rev: B
Table 8-4: Activity Matrix for Calibration Procedures
Calibration
Activities
Acceptance Limits
Frequency and
Method o f
Measurement
Action if
Requirements are
not Met
Zero-air
Zero-air, free of
contaminants
(Sec. 2.0.7 Ref. 11.).
Compare the new
Zero-air against
Source known to be
free of contaminants.
Take corrective action
with generation system
as appropriate.
Calibrator
Meet all requirement for
UV photometer as
specified in Sec. 2.7.2
QA Manual, TAD
2
and the
Fed. Reg.
1
or approve
Transfer Standard
Sec. 2.7.1, Q.A. Manual
and TAD
3
.
Re-certify transfer
Standard against
Primary UV
Photometer at least
Twice each quarter.
Return to supplier, or
take corrective action
with system as
appropriate.
Multipoint
According to Calibration
procedure (Sec. 2.7.2
Q.A.. Manual Ref 11) and
Federal Register; data
recorded.
Calibrate at least
Once, quarterly;
Anytime an audit
Indicates
discrepancy; After
maintenance that May
affect the Calibration
(Subsec 2.1) Federal
Register
1
.
Repeat the calibration.