Model 400E Ozone Analyzer
Instruction Manual
Getting Started
P/N 04316 Rev: B
39
3.3.2.
Basic Calibration Procedure
While it is possible to perform the following procedure with any range setting we
recommend that you perform this initial checkout using the 500 ppb range.
NOTE
The following procedure assumes that the instrument DOES NOT
have any of the available Zero/Span Valve Options installed and
Cal gas will be supplied through the Sample port.
See Section 5.3 for information regarding setting up of
Z/S Valve options.
See Section 7 for instructions for calibrating instruments
possessing Z/S Valve Options.
1.
Set the Analog Output Range of the M400E
SAMPLE
RANGE = 500.0 PPB
O3 =
< TST TST > CAL
SETUP
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
Press this button to set
the analyzer for
SNGL
DUAL
or
AUTO
range
(see Section 6.4)
SETUP X.X
RANGE: 500.0 CONC
0 0
5
0 0 .0
ENTR EXIT
To change the value of the Range
Setting, enter the number
sequence by pressing the key
under each digit until the expected
value appears.
SETUP X.X
RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Press this butto n to select
the Measurement Range
unit Type
(see Section 6.4.6)
EXIT
ignores the new setting and returns to the
previous menu.
ENTR
accepts the new setting and returns to the
previous menu..